Questions Some frequently asked questions - answered by Mike and others
|
Table of Contents
Printed spectrum is too small and on a corner of the paper. The following command will help you adjust for the size of the spectrum printed. Please refer to the following manual for this and much more information. Getting Started Chapter 9 , Display, Plotting and Printing Display Limits wc= width of chart. Adjust the value of this parameter . This parameter varies with the type of printer installed. To check the max value allow for the printer type wcmax? and set wc= wcmax. Also you can use the following commands to expand the width of you displayed spectrum. full for full width f for full spectrum range. Viewing two spectra simultaneously
NOTE: Prior to using addi, you should be sure you don't have any valuable unsaved spectra, parameters in exp5, because exp5 is the workspace used for this program. If you have never used exp5, you should be fine. To view two spectra at a time and see the difference between them: I. In VNMR, type: jexp1 and hit Return Printing Stacked spectra I. Load the spectra in exp1 and add them to an array: type ‘jexp1’ - join experiment 1 What do I do if the lock window disappears? If the lock window disappears try the following: a. Type acqi. If this doesn't work then try the following: a. exit VNMR b. logout of the console c. log back into the console and reopen VNMR If the window still does not appear then call someone on the NMR staff. The instrument needs to be rebooted. Back to TopWhat do I do if I break a sample in a magnet or in the NMR facility? There are two cases of broken samples. First, identify which category you fit into and then follow the directions given below: a. Broken NMR tube on the bench (i.e. not in the magnet) Then, you need to clean up the mess! Please wipe up all chemicals, remove broken glass from the NMR Facility, and take care of any other mess. If you broke a standard sample then please tell an NMR Staff Member so that the sample may be replaced. If it is after hours then leave a note in the logbook. b. Broken NMR tube in the magnet (i.e. the broken tube and its contents are now inside the NMR) Then, you need to notify an NMR Staff Member as soon as possible. The NMR Staff Member will need to know what was in the sample and what instrument you are on. If it is after hours then page someone on the contact list. YOU MUST CONTACT SOMEONE!!! If you fail to contact anyone and damage occurs to the probe, you are ultimately responsible. Back to TopWhy can't I log on as biopack? You obviously don't have the password Back to Top Why is the plotter undefined? The reason is that when your account was created default parameters were used. This means that your account has the wrong plotter configurations. On the inova300 the nearest plotter is called hplj4000 (landscape) and hplj4000_p(postcript). On the vnmr command line please type plotter='hplj4000' or plotter='hplj4000_p', or printer='lp'. Alternatively, you can use the menu buttons. [main]->[more]-> [configure]->[select plotter] Note new account are now created with the correct plotter and printers names. Back to Top When I have a spectrum I can only phase it in one small display at a time. You can only phase one small increment of your spectrum at a time because the 'phasing variable is set incorrectly. This is an old parameter present since the time when computers had limited memory capacity (e.g. Gemini's, XL...). On the vnmr command line consult the variable phasing? phasing=20 'this means 20% of your spectrum phasing=100 'this means 100% of your spectrum' Set phasing=100 Problem solved!!! Back to Top How to paste spectra by vnmr into Word or Word Perfect? Download the following document. Pasting_vnmr.pdf Back to Top Adding data files (FIDS) after a you stop(sa or aa) and experiment and was unable to resume (ra). When stopping an acquisition you can resume the acquisition with the command ra. Sometime the acquisition will restart from the last ct point but some other times it will not. You could save this experiment and begin another one and at the end add both data sets. Here is how to do it. Assuming you upload both experiments in exp1 and exp2 respectively clradd jexp1 add jexp2 add jexp5 setvalue('nt',<nt + extra_nt>,'processed') setvalue('ct',<ct + extra_ct>,'processed') wft
Provided by John Tomaszewski in an AMMRL letter to a user. Back to Top Printed spectrum is too small and in a corner of the paper. Part I The following command will help you adjust for the size of the spectrum printed. Please refer to the following VNMR manual for this and much more information about displaying spectrum. Getting Started Chapter 9 , Display, Plotting and Printing Display Limits wc= width of chart. Adjust the value of this parameter . This parameter varies with the type of printer installed. To check the max value allow for a given printer type wcmax? and set wc= wcmax. Also you can use the following commands to expand the width of you displayed spectrum. full =for full width f =for full spectrum range. Part II If this doesn;t work do the following. I have found that some of the accounts have wcmax set to 200 not 250. The maximum wcmax for laser printer is 250. Somehow wcmax have been reset by a previous user to 200 and it remains at 200 until someone for the same group resets it to 250. Set wcmax =250 then type wc=wcmax This should take care of the problem. Back to Top Unable to write, unable to jump experiments or display experimental library. One possible reason is, you are not a member of the user group. A staff member should check the /etc/group file and add your name to a group. another possible reason is that the ownership of the ownership of the experiment is not set up properly. the correct mode for the file under ~/vnmrsys/exp# should be rwerwer-e or 775. this is setup upon log in by a parameter know as umask which should be 002 i the files .login. Alos can be set up at the global level from the file ./etc/defaults/login Back to Top Rebooting the SUN If you have to reboot the SUN, do the following. First type su acqproc on the sun. This stops communication between the spectrometer console and the Sun computer. Then login as superuser: su see me for password then type: init 5 this will shut the computer down. To restart just push the button on the form of the sun computer Back to Top Rebooting spectrometer If you have to reboot the spectrometer, do the following. First type su acqproc on the sun. This stops communication between the spectrometer console and the Sun computer. You can now either reset the console cpu or completely shutdown the system. To reset the cpu: -Open the first cabinet door on the console -Press the reset button on the cpu board -wait for ~ 1 minute until the lights ppear on the remote display -type su acqproc again in the console For a complete console reboot:, after su acqproc -Go to the second console cabinet -At the bottom is a card cages switch. Turn this off (pull it towards you and down) -Then go to the back of the 2nd cabinet and flip the black switch down To start up the system: -flip the black switch at the back of the 2nd cabinet up -wait for 30 seconds until the fans start, then press the green button just above the black switch -after 2 minutes, go to the front of the second cabinet and turn the card cages switch on (pull towards you and up). -once again you have to wait one or two minutes until the lights on the remote display come up. The type su acqproc -Note after rebooting you will have to reset the temperature. In most cases you will have to search for the lock again, and the loack phase has a 75% chance of not being correct any longer Back to Top sdfsdfsfsdf |
|
|